TP1020 Automatic Tissue Processor
Automated tissue processing for histopathology labs.
| Type | Tissue Processor |
|---|
The Leica TP1020 is a carousel-type semi-enclosed benchtop tissue processor for histopathology laboratories, carrying tissue through dehydration, clearing and paraffin infiltration on a programmed schedule.
The instrument has 12 stations. Positions 1 to 10 hold reagents in 1.8 L containers; stations 11 and 12 are heated, temperature-controlled wax baths, also 1.8 L, with a third wax bath available in place of station 10. Wax bath temperature is adjustable across 45–65 °C.
The standard basket holds 100 cassettes, and a second basket takes capacity to 200 — useful where workload varies between routine days and a heavy list, since the same instrument covers both without a second run.
Nine programmes can be stored, with infiltration time per station programmable up to 99 hours 59 minutes. That range matters because processing schedules are not universal: fatty and dense tissue need longer, small biopsies need shorter, and running everything on one schedule is a common cause of blocks that will not section cleanly.
A vacuum option is available and can be applied at any station in manual or automatic operation, with a maximum pressure difference of 500 hPa. Vacuum substantially improves infiltration and shortens the schedule needed to achieve it.
Reagent rotation discipline matters more than any specification here — exhausted reagents are the commonest cause of poor sections. Rose Medicare supplies and services histopathology equipment across Bihar and Jharkhand.