histopathology

Leica Water Bath HI1220

Brand: Leica Availability: In stock

Precision water bath for histology slide preparation.

TypeWater Bath

The Leica HI1220 is a microprocessor-controlled water bath for floating out paraffin sections during microtomy, holding a stable temperature across the working surface.

Temperature is controllable from room temperature to 75 degrees C, with an operating environment of +15 to +40 degrees C.

Temperature control is the whole point of the instrument. Sections floated on water that is too cool will not flatten and retain compression folds from the microtome; too warm and they overstretch, distorting the tissue and producing artefacts a pathologist may have to interpret around. The correct setting sits a few degrees below the melting point of the paraffin in use, which is why a controllable and stable bath matters rather than any warmed vessel.

Stability across the surface matters as much as the set point. A bath with cool corners produces sections that behave differently depending on where they land, which is difficult to attribute and easy to blame on the microtome or the block.

For a histopathology laboratory the water bath sits between microtomy and staining, and poor sections at this stage cannot be recovered later. It is a small instrument with disproportionate influence on the quality of the final slide.

Rose Medicare supplies and services histopathology equipment across Bihar and Jharkhand.